CU-600GS Total Organic Carbon Analyzer
Introduction:
Ultraviolet (UV) is used to oxidize the slow-flowing water sample, and the organic matter in the sample is oxidized to CO2. The determination of CO2 adopts the conductivity detection technology, and the total organic carbon of the water sample is obtained by changing the conductivity of the water sample before and after oxidation. value.
Features:
1. High detection accuracy and low detection limit;
2. Conductivity temperature compensation technology to improve the accuracy of detection;
3. No need for carrier gas and reagents, simple maintenance and low cost;
4. A separate system suitability test interface, the detection curve can be seen in real time, which is more intuitive;
5. Large data storage capacity, easy to query, and can be queried by day and printed by batch number;
6. The UV lamp life countdown is convenient for the replacement and maintenance of the UV lamp;
7. It has the function of alarm output exceeding the upper limit;
8. Optional automatic calibration package, one-click to complete automatic calibration and system applicability test;
9. Optional automatic sampler to achieve unattended, saving manpower and time.
Technical parameter:
TOC detection range: 0.1 μg/L~1500.0ug/L
Conductivity detection range: 0.055-6.000uS/cm
Detectable items: TOC, TIC, TC
TOC detection accuracy: 0.1 μg/
Accuracy error: ≤±4%
Repeatability error: ≤2%
Zero drift: ±2%/D
Range drift: ±2%/D
Response time: ≤10min
Sample temperature: 0~100℃
Injection method: peristaltic pump
Ambient temperature: 10-60℃, the temperature change is within ±5℃/d
Relative humidity: ≤90%
Power requirements: 220V AC≤±10% 50Hz
Power: 60W
Base size: 750px×500px×450px
Weight: 8kg