CU-600 total organic carbon analyzer
Introduction:
The CU-600 Total Organic Carbon Analyzer uses ultraviolet (UV) to oxidize the slow-flowing water sample, and the organic matter in the sample is oxidized to CO2. The determination of CO2 adopts the conductivity detection technology. to obtain the total organic carbon value of the water sample. Application scope: laboratory offline detection of water for injection and purified water in the pharmaceutical industry; TOC online monitoring of water systems in the pharmaceutical industry; ultrapure water preparation system and wafer process in the semiconductor industry; deionized water preparation process in power plants.
Features:
1. High precision, low detection limit
2. Touch screen design, human interface, simple operation is convenient
3. The conductivity temperature compensation technology, improve the accuracy of detection
4. No need for carrier gas and reagents, simple maintenance and low cost
5. Separate system applicability test interface, operating more human
6. Testing curve in real time, more intuitive
7. Large data storage, query convenient, and can query by date
8. The built-in stylus printer, reduce the occupied space
9. Uv lamp life countdown, convenient for uv lamp replacement and maintenance
10. External 4-20 ma
11. Beyond the limit alarm output function
12. Optional on-line sampling device, the realization of the water system of on-line monitoring
13. Optional automatic calibration package, a key to complete the automatic calibration and system applicability test
14. Optional automatic sampling instrument, to realize unattended, save manpower and time
Technical parameters
TOC detection range: 0.1 ~ g/L ~ 1500.0ug/L
TOC detection accuracy: 0.1 g/L (1.0uS/cm)
Can be detected items: TOC, electrical conductivity, temperature
Electrical conductivity detection range: 0.055-6.000uS/cm
TOC error: less than 5%
Repeatability error: less than 3%
Time: the analysis of the shortest 20s interval (set).
Response time: 15min
Sample temperature: 0 ~ 100
Zero drift: + 2%/D
Range drift: + 2%/D
Sample way: the pump
Environment temperature: the temperature change of 10-40 is within 5 /d.
Relative humidity: less than 85%
Power supply: 220V AC = 50Hz + 10%
Power: 60W
Basic size: 30cm x 20cmm x 18cm
Weight: 10kg